Article title

The effects of laser marking and symbol etching on the fatigue life of medical devices.

Type
D - Journal article
Title of journal
Journal of Medical Engineering
Article number
-
Volume number
2013
Issue number
570354
First page of article
1
ISSN of journal
2314-5129
Year of publication
2013
Number of additional authors
2
Additional information

Following in-vitro trials of medical devices it became apparent that laser marking was not a passive marking technology (as regards fatigue) as previously depicted. Laser marking is the most common form of permanent marking and as implantable devices are becoming ever more slight in nature its effects on fatigue are becoming more important. Following an extensive literature review we found that little had been published. We therefore conducted a targeted research study into the effects of laser marking on fatigue properties of medical devices. We know of one litigation case that is using this output as a basis for evidence.